Chromatography Research Today is a free monthly online journal that collates and summarizes the latest research about Chromatography, including details on column chromatography, gas chromatography (gc), liquid chromatograpy, hplc. | ||||||||
|
Quantitative analysis of Variolin analog (PM01218) in mouse and rat plasma by high-performance liquid chromatography/electrospray ionization tandem mass spectrometry.Yin J, Aviles P, Ly C, Lee W, Guillen MJ, Munt S, Cuevas C, Faircloth G PharmaMar USA Inc., 320 Putnam Avenue, Cambridge, MA 02139-4616, USA. jyin@pharmamarusa.com PM01218 is a novel marine-derived alkaloid and has shown potent growth inhibitory activity against several human cancer cell lines. A rapid and sensitive high performance liquid chromatography/tandem mass spectrometry (HPLC-MS/MS) method was developed and validated to quantify PM01218 in mouse and rat plasma. The lower limit of quantitation (LLOQ) was 0.05 ng/mL. The calibration curve was linear from 0.05 to 100 ng/mL (R(2)>0.999). The assay was specifically based on the multiple reaction monitoring (MRM) transitions at m/z 278.4-->184.2, no endogenous material interfaced with the analysis of PM01218 and its internal standard from blank mouse and rat plasma. The mean intra- and inter-day assay accuracy remained below 15 and 8%, respectively, for all calibration standards and QC samples. The intra- and inter-day assay precision was less than 12.8 and 8.5% for all QC levels, respectively. The utility of the assay was demonstrated by pharmacokinetics studies of i.v. (bolus) PM01218 on SD rats. Published 7 March 2006 in J Chromatogr B Analyt Technol Biomed Life Sci, 832(2): 268-73.